MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number 97715
Device Problems Failure to Deliver Energy (1211); Data Problem (3196)
Patient Problem Therapeutic Effects, Unexpected (2099)
Event Date 07/07/2020
Event Type  Malfunction  
Manufacturer Narrative

If information is provided in the future, a supplemental report will be issued.

 
Event Description

Information was received from a patient regarding their implantable neurostimulator (ins). Information was reported that the patient stated he was charging the ins and the controller went blank/unresponsive. The patient stated that in additional he is not getting any stimulation. The caller stated it happened before that the controller went blank, but the ins was still functioning. The patient stated now it seems like neither are working. The patient stated he tried resetting the controller four times last night and that did not resolve the issue. The patient was walked through resetting the controller and it powered on and the device showed the device was in mri mode. The caller exited mri mode and turned stimulation back on. The patient stated everything is working now. The controller battery is at 50% and the ins is at 90%. No further complications were reported. No patient symptoms were reported.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameINTELLIS
Type of DeviceSTIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key10252408
MDR Text Key198169211
Report Number3004209178-2020-11894
Device Sequence Number1
Product Code LGW
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberP840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type CONSUMER
Reporter Occupation
Type of Report Initial,Followup
Report Date 08/19/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received07/09/2020
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device EXPIRATION Date05/28/2020
Device MODEL Number97715
Device Catalogue Number97715
Was Device Available For Evaluation? No
Is The Reporter A Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/12/2020
Was Device Evaluated By Manufacturer? No Answer Provided
Date Device Manufactured06/06/2019
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-