• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 500; MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 500; MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500 Back to Search Results
Model Number 500AVHCT
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Per the user facility's biomedical technician, the issue occurred about two weeks before it was reported.The inaccurate values were for partial pressure of carbon dioxide (pco2).
 
Event or Problem Description
It was reported that prior to the use of the device for a cardiopulmonary bypass (cpb) procedure, the blood parameter monitor (bpm) values were inaccurate.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Additional Manufacturer Narrative
The reported complaint could not be confirmed.During laboratory analysis, the product surveillance technician (pst) observed that the on-screen values could be adjusted to sample values throughout the evaluation.A sample 540 calibration was completed successfully.The monitor operated as intended throughout the evaluation.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event or Problem Description
Per clinical review: multiple diligence attempts were unsuccessful in obtaining additional clinical information.The details available are that the values were off from what was expected when the monitor was being used during an extracorporeal membrane oxygenation (ecmo) procedure.
 
Additional Manufacturer Narrative
H3: 81 - evaluation is in progress, but not yet concluded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CDI BLOOD PARAMETER MONITORING SYSTEM 500
Common Device Name
MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key10253428
Report Number1828100-2020-00256
Device Sequence Number18961608
Product Code DRY
UDI-Device Identifier00886799001646
UDI-Public(01)00886799001646(11)990901
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K133658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number500AVHCT
Device Catalogue Number500AVHCT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/23/2020
Supplement Date Received by Manufacturer07/16/2020
08/13/2020
Initial Report FDA Received Date07/09/2020
Supplement Report FDA Received Date08/05/2020
08/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
1828100-08/07/15-002-C
Patient Sequence Number1
-
-