• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS; TORIC IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS; TORIC IOLS Back to Search Results
Model Number ZCT600
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Vitrectomy (2643); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: exact date unknown/asked but unavailable.Best estimate is between (b)(6) 2019 - (b)(6) 2020.(b)(4).The device was manufactured at the (b)(4) site which has been closed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the intraocular lens (iol) was explanted from the patient's left eye due to the patient having a history of keratoconus, therefore, it is difficult to get accurate measurements.The lens was not the right fit for the patient after all.The patient had blurry vision and had a hard time reading.The incision was enlarged, unplanned sutures, and an anterior vitrectomy were required.The replacement lens was successfully implanted with a different model and lower diopter.No additional information was provided.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Additional Manufacturer Narrative
Additional information: section d10: device available for evaluation? yes.Returned to manufacturer on: 7/13/2020.Section h3: device returned to manufacturer ¿ yes.Device evaluation: the complaint lens was received in a specimen cup.Visual inspection under magnification revealed viscoelastic residue on the optic body and haptics, and that the lens was received cut in half, which is consistent with a lens that was handled during explant.The lens was also observed to be returned covered in fibers, which may have occurred during the handling of the lens after explant, and therefore cannot be confirmed to be related to manufacturing.Based on the return condition of the lens no further product evaluation could be performed.The complaint issue could not be confirmed, and no product deficiency was identified.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured according to specifications.The search revealed that no similar complaint for this production order number has been received.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.As a result of the investigation, with limited information available, it could not be determined if there.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key10253878
Report Number2020664-2020-00038
Device Sequence Number19502786
Product Code HQL
UDI-Device Identifier05050474604025
UDI-Public(01)05050474604025(17)220620
Combination Product (Y/N)Y
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/29/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/20/2022
Device Model NumberZCT600
Device Catalogue NumberZCT600U130
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2020
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 06/12/2020
Supplement Date Received by Manufacturer08/05/2020
10/25/2020
Initial Report FDA Received Date07/09/2020
Supplement Report FDA Received Date08/29/2020
10/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-