|
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device has damage and debris within the 2.5mm guide, rendering the device inoperative.The tip of the guide is also damaged.The device shows signs of extensive use.The clinical/medical evaluation concluded that based on the information provided, the surgeon used another 2.5 drill guide assembled with other components of the system to complete the procedure.Although, this happened inside of the patient, per report, nothing remained inside of the patient.Since there was no delay or patient injury/harm, no further clinical/medical assessment is warranted at this time.Should additional medical information be provided, this complaint will be re-assessed.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|