| Model Number |
97014 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Swelling (2091); Reaction (2414); No Code Available (3191)
|
| Date of Event |
06/13/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The exact lot number involved in this event is unknown.The device history record for potential lots 30060846, 30057604, and 30063592 were reviewed and the products were produced according to product specifications.The sample is reported to be available, but has not yet been received by the manufacturer.All information reasonably known as of 09 jul 2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical inc.Complaint database and identified as complaint (b)(4).
|
| |
|
Event or Problem Description
|
|
Avanos medical inc.Received a single report that referenced seven different incidences, which were associated with separate units, involving seven different patients.This is the fourth of seven reports.Refer to 3006646024-2020-00018 for the first report, refer to 3006646024-2020-00019 for the second report, refer to 3006646024-2020-00020 for the third report, refer to 3006646024-2020-00022 for the fifth report, refer to 3006646024-2020-00023 for the sixth report, refer to 3006646024-2020-00024 for the seventh report.It was reported that a newly intubated patient developed sudden angioedema following mouth care.The patient's eyes were swollen shut, cheeks and lips were swollen, and "tongue is out of her mouth." additional information was received 07-jul-2020 in user facility submitted medwatch report mw5095069.The event description stated, "used avanos oral care 24 hour kit (q4 hour) 97014 on 4 patients and each developed swelling of the lips, cheeks, and tongue.Fda safety report id# [number redacted].".
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: d10.The actual complaint product used on the patient was not available for evaluation.Representative samples from each possible lot were tested and met specification for appearance, ph, and assay.Root cause could not be determined.All information reasonably known as of 17 sep 2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
| |
|
Event or Problem Description
|
|
Additional information received 09-jul-2020 indicated the patient was admitted on (b)(6) 2020.Patient was intubated (b)(6) 2020 and mouth care kit was used at 1224.On (b)(6) 2020 at 1730 face, mouth, and lips edema was noted.Benadryl 50mg was given every 6 hours.Oral care kit usage was discontinued (b)(6) 2020 at 1701.Patient was safely extubated and ultimately discharged to a nursing facility.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: b3.Additional information: a2, a4, b5, b7.All information reasonably known as of 03 aug 2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
| |
|
Search Alerts/Recalls
|