Model Number 928850
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/24/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Device expiration date: unknown.Device manufacture date: unknown.Investigation summary: level a investigation - complaint evaluation / complaint history check for the event(s) that occurred.Severity: s_1__; occurrence: unable to perform complaint lot history check for needle damaged/hole due to unknown lot number.A review of risk management 150rmn-0001-16 revision 13 indicates that the potential risk of this specific reported incident (syringe, needle damaged) was captured and addressed appropriately investigation summary: no samples were returned therefore the complaint could not be confirmed and the root cause is undetermined.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Unable to perform dhr check for needle damaged/hole due to unknown lot number.Investigation conclusion: as no samples and/or photo(s) were received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure as no samples or photos were returned.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.Root cause description: root cause cannot be determined at this time as the issue is unconfirmed as no samples or photos were returned.Rationale: based on the investigation, no additional investigation and no capa is required at this time.
 
Event or Problem Description
It was reported that the syringe 1.0ml 29ga 1/2in 10bag 500 pl/wg experienced a failure to contain medication.It has not been specified whether the product defect was noted prior to, during, or after use.The following information was provided by the initial reporter: material no.928850 batch no.Unknown.Consumer left voicemail stating that there is a hole in the middle of the needle.Date of event: unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE 1.0ML 29GA 1/2IN 10BAG 500 PL/WG
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
MDR Report Key10255671
Report Number1920898-2020-00831
Device Sequence Number11386763
Product Code FMF
UDI-Device Identifier00311917014876
UDI-Public00311917014876
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K170386
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number928850
Device Catalogue Number928850
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2020
Initial Report FDA Received Date07/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-