MAUDE Adverse Event Report: ARTHROCARE CORP. SUPER TURBOVAC 90 IFS; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. SUPER TURBOVAC 90 IFS; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number ASH4250-01
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Information (3190)
Date of Event 12/24/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
It was reported that the wand did not connect; the rf unit did not recognize the wand.It is unknown whether the event happened during surgery and if there was patient involvement and if there was a back-up device available or if there was a delay.No further details were provided.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H3, h6: the device, intended for use in treatment, was not returned for evaluation.A relationship, if any, between the subject device and the reported event could not be determined since the product was not returned for evaluation.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.Visual inspection and functional testing could not be performed since the device was not returned for evaluation.Thus, the complaint could not be verified, nor could a root cause be determined with confidence.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.Potential factors unrelated to the manufacture or design of the device that could have contributed to the reported event includes: v.(1) previous use (2) disconnecting the wand from the controller after power has been turned on.(3) an issue with one of the concomitant devices in use at the time of this complaint event.There are no indications to suggest the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPER TURBOVAC 90 IFS
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key10257330
Report Number3006524618-2020-00461
Device Sequence Number3458484
Product Code GEI
UDI-Device Identifier00817470006506
UDI-Public00817470006506
Combination Product (Y/N)N
Initial Reporter CountrySI
PMA/510(K) Number
K083306
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberASH4250-01
Device Catalogue NumberASH4250-01
Device Lot Number2037839
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/24/2019
Supplement Date Received by Manufacturer07/29/2020
Initial Report FDA Received Date07/10/2020
Supplement Report FDA Received Date08/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-