• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC SMALL BATTERY DRIVE II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC SMALL BATTERY DRIVE II; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.110
Device Problem Excessive Heating (4030)
Patient Problem Not Applicable (3189)
Event Date 06/29/2020
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Concomitant medical products and therapy dates: small battery drive casing device, (b)(6) 2020.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If additional information should become available, a supplemental medwatch report will be submitted accordingly.(b)(4).
 
Event Description
It was reported that during pre-surgery, it was discovered that the small battery drive device overheated while being used together with the battery drive casing device.It was further reported that the device was spinning slowly.It was not reported if there were any delays in the surgical procedure or if a spare device was available for use.There was no human patient involvement as the device was a veterinary device.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that there was an unknown substance on the motor and motor plate.It was further observed that the device had a corrosive substance and was overheating.It was further determined that the device would not run.It was further determined that the device failed pretest for check oscillation/forward/reverse mode function assessment and check power with test bench assessment ¿ device not running/functioning.Therefore, the reported condition was confirmed.The assignable root cause was determined to be traced to maintenance, which is improper maintenance.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMALL BATTERY DRIVE II
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key10258034
MDR Text Key198345294
Report Number8030965-2020-04666
Device Sequence Number1
Product Code HWE
UDI-Device Identifier07611819408357
UDI-Public07611819408357
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 06/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.110
Device Lot Number1399
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/08/2020
Initial Date Manufacturer Received 06/29/2020
Initial Date FDA Received07/10/2020
Supplement Dates Manufacturer Received07/23/2020
Supplement Dates FDA Received08/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-