• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN TEMPERATURE MANAGEMENT SYSTEM, ARCTICGEL PADS ARCTIC GEL PADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN TEMPERATURE MANAGEMENT SYSTEM, ARCTICGEL PADS ARCTIC GEL PADS Back to Search Results
Catalog Number 318-02
Device Problems Defective Device (2588); Misassembly by Users (3133)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that neonate started to rewarm around 530pm. Rewarm should have been complete at 1130pm but the timer reads 1 hour 40 minutes remaining. The baby was up to 35. 6c at one time, but was down to 35. 4c now in arctic sun device serial no: (b)(4). Patient was 35. 5c. Rewarm tab read to rewarm from 36c to 36c at 0. 4c/hr. Nurse confirms their protocol was to rewarm to 36c, she was not sure of rate but assumed 0. 4c/hr is correct. Patient was currently 35. 5c. Water 36. 6c. Flow 0. 8lpm. Ip -7psi, cp 27%. Nurse disconnected and reconnected the pad rotating the orientation of the clamp. Flow now was 1. 1lpm. The straps were not around the abdomen. Mss asked nurse to secure the straps and "taco" the baby and mss suggested turning on radiant heat if their protocol allows and these measures are not successful. Trend was neutral at beginning of call, now pointing up. Temperature up to 35. 6c. Confirmed the timer is counting up, not down. As per the follow up information, the nurse stated therapy completed and pads disposed of after use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN TEMPERATURE MANAGEMENT SYSTEM, ARCTICGEL PADS
Type of DeviceARCTIC GEL PADS
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key10258299
MDR Text Key200613951
Report Number1018233-2020-04460
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142702
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/17/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number318-02
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/26/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-