• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX M150; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX M150; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 115307
Device Problems Fluid/Blood Leak (1250); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Patient Involvement (2645)
Event Date 06/17/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during priming with three prismaflex m150 sets, external fluid leakage from the effluent bags were observed.It was reported no alarm was triggered.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The device was received for evaluation.A leak test was performed and a leak at the level of the drain valve.Was observed.During filling, the drain valve finally gave out.The reported condition was verified.The cause was misuse of the bag.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.It is to be noted that the prismaflex m150 set ckt and all the accessories provided within the set, including the drain bag, is a single use product.If the bag was in use for several hours, the bag had been filled and emptied several times.The filling/emptying cycles generates physical constraints on the bags.The prismaflex control unit operator's manual indicates to ¿change fluid bags when the appropriate caution alarm occurs (pbp bag empty, replacement bag empty, dialysate bag empty or effluent bag full)¿.The prismaflex control unit warns the operator when the effluent bag is full, and it is specified in the on-screen instruction delivered by the prismaflex control unit that the drain bag should be disconnected and changed for a new bag.Only one bag is provided within the sets since this bag is the one needed to prime the set and to start therapy.Since therapy duration and prescribed doses vary for each patient and therapy, replacement bags are provided by baxter as spare bags.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX M150
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10258607
MDR Text Key198360671
Report Number8010182-2020-00085
Device Sequence Number1
Product Code KDI
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number115307
Device Lot Number20C1205
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/31/2020
Initial Date Manufacturer Received 06/17/2020
Initial Date FDA Received07/10/2020
Supplement Dates Manufacturer Received08/06/2020
Supplement Dates FDA Received08/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
NI.
-
-