• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I NANOTITE TM TAPERED CERTAIN PREVAIL IMPLANT 5/4 X 15MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I NANOTITE TM TAPERED CERTAIN PREVAIL IMPLANT 5/4 X 15MM; DENTAL IMPLANT Back to Search Results
Model Number NIITP5415
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Inflammation (1932); Pain (1994); No Code Available (3191)
Date of Event 03/10/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet (b)(4).A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
Doctor reports that patient was seen for an emergency exam due to complaint of pain at niitp5415 implant #27 site.Radiograph revealed peri implant radiolucency.Implant was mobile.Implant was removed and site was grafted using puros bone graft material.Inflammation is also reported.
 
Additional Manufacturer Narrative
This report is being submitted to relay additional information.The following sections are being reported: b4: date of this report was updated.D4: unique identifier (udi) number was updated.D10: device availability was updated.G4: date received by manufacturer was updated.G7: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H4: device manufacturer date was updated.H6: method code was updated.H10: narrative/data was updated.The reported device was received for evaluation.The reported medical condition of peri implantitis was not confirmed.Visual inspection was not performed.Functional testing to recreate the reported event could not be performed due to the nature of the device and event.A device history record (dhr) review was completed for the reported device lot.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformance, which could have caused or contributed to the reported event was noted as part of the dhr.A sterilization record review was completed and all sterilization activities were conducted with no nonconformities.A complaint history review was completed for the reported device lot for similar event and no other complaint was identified.A definitive root cause for the reported event could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional or corrected information to report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NANOTITE TM TAPERED CERTAIN PREVAIL IMPLANT 5/4 X 15MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key10259786
Report Number0001038806-2020-00966
Device Sequence Number11315772
Product Code DZE
UDI-Device Identifier00844868011002
UDI-Public(01)00844868011002(17)190728(10)2014070790(241)NIITP5415
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
K063341
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 08/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date07/28/2019
Device Model NumberNIITP5415
Device Catalogue NumberNIITP5415
Device Lot Number2014070790
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/08/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/24/2020
Supplement Date Received by Manufacturer07/15/2020
Initial Report FDA Received Date07/10/2020
Supplement Report FDA Received Date08/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient Weight108
-
-