The patient's age at time of event or dob, gender, and weight are unknown.This information was not available from the facility.During inflation, the balloon ruptured above rbp.Recurrence of this malfunction could result in a flow limiting dissection, embolism, or perforation.No patient injury reported.Cross reference mfr report numbers: 3005462046-2020-00014.Patient information regarding relevant tests/laboratory data or medical history are unknown.This information was not available from the facility.Report source: foreign- (b)(6).The angiosculpt device was infectious and discarded by the facility, thus no returned product investigation was performed.The angiosculpt device was used off-label.The ifu states to inflate the balloon 2 atm every 10-15 seconds until full balloon inflation is achieved and to not exceed the rbp printed on the package label.
|