It was reported that left hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the bhr cup and hemi head was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the head and sleeve.Similar complaints have been identified for the cup.However, as the device is no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.Review of manufacturing records did not reveal any waivers, concessions, manufacturing or material abnormalities that could have contributed to this issue.It was also confirmed that all devices were sterilised.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event.The available medical documents were reviewed.The reported elevated cobalt and chromium and intraoperative findings of milky and stained gray synovium may be consistent with findings associated with metallosis; however, the root cause of the elevated metal ions, milky and gray stained synovium and metallosis cannot be confirmed and it cannot be concluded that the reported reactions were associated with a mal-performance of the implant.No post revision medical documentation was provided.The patient impact beyond the pain, revision, and expected transient postop convalescence period cannot be determined.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
|