SMITH & NEPHEW MEDICAL LTD. UNKN PICO 7; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Catalog Number UNKN05004902
Device Problem Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/16/2020
Event Type  malfunction  
Event Description
It was reported that when using two pico 7 devices, all lights illuminated and stayed on.No information about delay.The treatment continued with the same device.No patient harm reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
We have now completed our investigation into the reported complaint.A review of the batch manufacturing records could not be performed as no part or lot numbers were provided, however, there are no indications to suggest that the device did not meet specifications upon release into distribution.A complaint history review was carried out across all pico 7 products, there have been further complaints reported with this failure mode in the past three years.As no product could be returned, a thorough evaluation of the device could not be carried out.The devices were used for treatment.It was reported that all indicator lights were solidly illuminated.This means that the devices entered a non-recoverable error state.There are various reasons that the device may enter an error state, such as out of range negative pressure, out of range power supply and an error retrieving data.If the pump does enter an error state, it must be replaced.This is a patient safety feature.However, it was also reported that treatment was continued with the same device so it is not probable that all lights were illuminated.It is likely that the leak, battery low or dressing full indicators may have been illuminated to notify of a potential issue.We have not been able to confirm a relationship between the event and the devices or identify a definitive root cause.No further actions by smith and nephew are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.Smith and nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN PICO 7
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key10261971
MDR Text Key199064021
Report Number8043484-2020-01852
Device Sequence Number1
Product Code OMP
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 08/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKN05004902
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/19/2020
Initial Date FDA Received07/12/2020
Supplement Dates Manufacturer Received08/07/2020
Supplement Dates FDA Received08/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-