MAUDE Adverse Event Report: WARSAW ORTHOPEDICS METRX SYSTEM; LIGHT, SURGICAL, CARRIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WARSAW ORTHOPEDICS METRX SYSTEM; LIGHT, SURGICAL, CARRIER Back to Search Results
Model Number 9560757
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/09/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from health care professional via manufacturing representative regarding patient with stenosis spondylolisthesis for spinal therapy.Patient demographics: gender: male; age: (b)(6).Outcome: alive no injury.Event was intra-op.Procedure used was mast tlif.Levels implanted: l4-s1 it was reported that the light source is inserted into the metrx tube per its ifu.However, after about an hour, the light source comes apart at the end.The two pieces of metal basically split in two, with one piece falling off, exposing the light.Of course the light is hot so this poses a burn risk to the patient as well as the surgical staff.It also poses a fire hazard to the patient and the or.Lastly, there is the potential for the one piece of metal that falls off to become lodged into the patient or fall into the would be unknown to the surgeon and staff.Fortunately in this case, none of these more serious issues occurred.The lights and the piece of disassociated metal were removed from the patient and discarded.No patient symptoms were reported.No further complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METRX SYSTEM
Common Device Name
LIGHT, SURGICAL, CARRIER
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key10262744
Report Number1030489-2020-00877
Device Sequence Number14318427
Product Code FSZ
UDI-Device Identifier00721902663444
UDI-Public00721902663444
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K993021
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/13/2024
Device Model Number9560757
Device Catalogue Number9560757
Device Lot Number0715120W
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/09/2020
Initial Report FDA Received Date07/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age35 YR
-
-