| Model Number |
MMT-523RNAS |
| Medical Device Problem Codes |
Device Difficult to Program or Calibrate (1496); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
06/29/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer's daughter reported via phone call that customer experienced high blood glucose.The customer¿s blood glucose level was over 600 mg/dl.Customer was treated with insulin pump.The customer reported that act and escape buttons were not responding, buttons were getting harder.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with intermittent button response due to flattened button dome switch and no unlocked j2/lcd keypad connector noted during visual inspect.Unable to perform all functional testing including the displacement, operating currents, a21 error test, rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery test due to buttons not responding.
|
| |
|