• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523NAH
Medical Device Problem Codes Battery Problem (2885); Power Problem (3010)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/06/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event or Problem Description
Information received by medtronic indicated that the insulin pump had battery failed error.The customer stated insulin pump had battery lasting for 7 days only.No harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
 
Additional Manufacturer Narrative
Device passed the operating currents, self test and displacement test.No charge or battery less than expected anomaly and no failed battery test alarm noted during test.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAH
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key10263024
Report Number2032227-2020-145395
Device Sequence Number2345441
Product Code OYC
UDI-Device Identifier00643169507173
UDI-Public(01)00643169507173
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523NAH
Device Catalogue NumberMMT-523NAH
Device Lot NumberB4523NAHJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2020
Initial Date Received by Manufacturer 07/07/2020
Supplement Date Received by Manufacturer09/30/2020
Initial Report FDA Received Date07/13/2020
Supplement Report FDA Received Date10/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age16 YR
Patient Weight93
-
-