ENCORE MEDICAL L.P. LINEAR HIP; LINEAR, STD, REDUCED NECK, SZ 10, W/P2 COATING Back to Search Results
Model Number 425-98-010
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 06/15/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to these documents as soon as it becomes available.
 
Event or Problem Description
Revision surgery - the patient had a bone fracture.
 
Additional Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was reported as fracture.The previous surgery and the surgery detailed in this event occurred 27 days apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to fracture.There were no findings during this evaluation that indicate the reported devices were defective.No other information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.Due to the short time between previous and revision surgery, it is also possible that the event may have been occurred due to lack of post-operative care or patient noncompliance with medical instructions.There are multiple factors that may also contribute to the event that are outside the control of djo surgical such as poor bone density, inadequate soft tissue support, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR HIP
Common Device Name
LINEAR, STD, REDUCED NECK, SZ 10, W/P2 COATING
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445,
MDR Report Key10264292
Report Number1644408-2020-00584
Device Sequence Number3463418
Product Code LPH
UDI-Device Identifier00888912078825
UDI-Public(01)00888912078825
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K120241
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/18/2025
Device Model Number425-98-010
Device Catalogue Number425-98-010
Device Lot Number562G1122
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/15/2020
Supplement Date Received by Manufacturer07/23/2020
Initial Report FDA Received Date07/13/2020
Supplement Report FDA Received Date08/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
400-03-402 LOT 871B1228; 400-03-402 LOT 871B1228
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
-
-