• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I NANOTITE TM TAPERED CERTAIN PREVAIL IMPLANT 5/4 X 10MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I NANOTITE TM TAPERED CERTAIN PREVAIL IMPLANT 5/4 X 10MM; DENTAL IMPLANT Back to Search Results
Model Number NIITP5410
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Edema (1820); No Code Available (3191)
Date of Event 10/19/2012
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Weight unknown / not provided.A summary investigation has been completed for bone loss events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
It was reported that patient had bone loss.Edema.At tooth location 31.Implant was removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NANOTITE TM TAPERED CERTAIN PREVAIL IMPLANT 5/4 X 10MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key10264595
Report Number0001038806-2020-00973
Device Sequence Number2329610
Product Code DZE
UDI-Device Identifier00844868010975
UDI-Public(01)00844868010975(17)160401(10)1018470(241)NIITP5410
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K063341
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2020
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date04/01/2016
Device Model NumberNIITP5410
Device Catalogue NumberNIITP5410
Device Lot Number1018470
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/29/2020
Initial Report FDA Received Date07/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
-
-