| Catalog Number |
5552002 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
06/22/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned for evaluation.Visual inspection demonstrates a hair inside of the sealed product packaging.The foreign material does not extend beyond the seal which remains intact.Based on the investigation performed and sample evaluation, the foreign hair presented during the manufacturing process prior to the packaging being sealed.To date, this is the only reported complaint for this manufacturing lot of (b)(4) units released for distribution in february, 2020.Awareness training was provided to the appropriate associates.
|
| |
|
Event or Problem Description
|
|
As reported, there was a hair partially inside the unopened package of an x-stream tubing set w/ nd trumpet valve.It is alleged that the unopened packaging was contaminated.The device was not used on the patient.
|
| |
|