PHYSIO-CONTROL, INC. - 3015876 LIFEPAK 15 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Medical Device Problem Code Failure to Deliver Energy (1211)
Health Effect - Clinical Codes No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/17/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Physio-control evaluated the customer's device and verified that the event code was logged in the device's memory; however, the issue could not be duplicated during testing.Physio cleared event codes from the device's memory.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.The cause of the reported issue could not be determined.
 
Event or Problem Description
The customer contacted physio-control to report that there was an event code logged in the device's memory.The event code logged is related to a potential failure of the device to deliver defibrillation energy.There was no report of patient use associated with the reported event.
 
Additional Manufacturer Narrative
Physio replaced the device¿s main keypad assembly.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.Based on the information available, physio-control has determined that the likely cause of the reported issue was that the shock switch located on the main keypad assembly was out of tolerance due to excessive resistance.When this issue occurs, an event code will be logged in the device's memory following a failure of the device to deliver defibrillation energy.
 
Event or Problem Description
The customer contacted physio-control to report that there was an event code logged in the device's memory.The event code logged is related to a potential failure of the device to deliver defibrillation energy.There was no report of patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK 15 DEFIBRILLATOR/MONITOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key10268208
Report Number0003015876-2020-00882
Device Sequence Number3043847
Product Code MKJ
UDI-Device Identifier00883873911556
UDI-Public00883873911556
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Remedial Action Notification
Type of Report Initial,Followup
Report Date (Section B) 10/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001263
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/17/2020
Initial Date Received by Manufacturer 06/17/2020
Supplement Date Received by Manufacturer09/22/2020
Initial Report FDA Received Date07/13/2020
Supplement Report FDA Received Date10/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3015876-11/18/2019-001C
Patient Sequence Number1
-
-