| Model Number |
MMT-523LNAH |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
07/04/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that they experienced high blood glucose.The customer's blood glucose was 440 mg/dl.The customer was treated with the manual injection, insulin pen.The customer been using the insulin pump system within 48 hours of reported high blood glucose event.The customer was neither in emergency room, nor admitted into hospital, as a result of high blood glucose.It was stated that the drive support cap was appeared normal.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime or a33 test, excessive no delivery test and delivery accuracy test at 0.08735 inches.The stop current and run current measurement tests are within specification.Device also passed self test, off no power alarm test and a21 error test.The no delivery alarm functions properly during the basic occlusion test and occlusion test.No unexpected no delivery alarm noted during testing.Device uploaded properly using carelink.Device had cracked battery tube threads, cracked case at the reservoir tube window corner and cracked reservoir tube lip.The test p-cap and reservoir does lock in place in the reservoir compartment.(b)(4).
|
| |
|
Search Alerts/Recalls
|