Model Number MMT-523LNAH
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 07/04/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that they experienced high blood glucose.The customer's blood glucose was 440 mg/dl.The customer was treated with the manual injection, insulin pen.The customer been using the insulin pump system within 48 hours of reported high blood glucose event.The customer was neither in emergency room, nor admitted into hospital, as a result of high blood glucose.It was stated that the drive support cap was appeared normal.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime or a33 test, excessive no delivery test and delivery accuracy test at 0.08735 inches.The stop current and run current measurement tests are within specification.Device also passed self test, off no power alarm test and a21 error test.The no delivery alarm functions properly during the basic occlusion test and occlusion test.No unexpected no delivery alarm noted during testing.Device uploaded properly using carelink.Device had cracked battery tube threads, cracked case at the reservoir tube window corner and cracked reservoir tube lip.The test p-cap and reservoir does lock in place in the reservoir compartment.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAH
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key10268878
Report Number2032227-2020-145683
Device Sequence Number13655932
Product Code OYC
UDI-Device Identifier00643169513839
UDI-Public(01)00643169513839
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-523LNAH
Device Catalogue NumberMMT-523LNAH
Device Lot NumberA4523LNAHJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2020
Initial Date Received by Manufacturer 07/07/2020
Supplement Date Received by Manufacturer08/10/2020
Initial Report FDA Received Date07/13/2020
Supplement Report FDA Received Date08/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FRN-MMT-326-RSVR, UNOMED SET
Outcome Attributed to Adverse Event Other;
Patient Age87 YR
Patient Weight54
-
-