| Model Number |
MMT-722LNAS |
| Medical Device Problem Code |
Priming Problem (4040)
|
| Health Effect - Clinical Codes |
Hyperglycemia (1905); Hypoglycemia (1912)
|
| Date of Event |
01/13/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump was not priming.The customer stated they had low blood glucose level and treated with food.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test, rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery test.Device primed properly.
|
| |
|