• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI INC CW-USLTA CYBERWAND ADVANCED TRANSDUCER; LITHOTRIPTOR, ELECTRO-HYDRAULIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI INC CW-USLTA CYBERWAND ADVANCED TRANSDUCER; LITHOTRIPTOR, ELECTRO-HYDRAULIC Back to Search Results
Model Number CW-USLTA
Device Problem Material Puncture/Hole (1504)
Patient Problem No Patient Involvement (2645)
Event Date 06/16/2020
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus and evaluated.The reported problem was confirmed.A conclusive root cause was not determined.
 
Event Description
A user facility reported that, upon receipt of the device, a hole was observed in the rubber of the power cord.There was no patient or end user injury or harm, associated with the problem reported to olympus.
 
Manufacturer Narrative
This supplemental report is submitted to provide the results of the legal manufacturer's investigation and device history record (dhr) review.This device was manufactured in may, 2019 by the oem.The dhr's were made available to olympus for review.All records indicate that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.A definitive root cause could not be determined, however, it is believed that the observed damage to the transducer cord was a cut created by a blade.This cut may have been incidental to clinical use or other general mishandling during unpacking or transportation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CW-USLTA CYBERWAND ADVANCED TRANSDUCER
Type of Device
LITHOTRIPTOR, ELECTRO-HYDRAULIC
Manufacturer (Section D)
GYRUS ACMI INC
136 turnpike road
southborough PA 01772
MDR Report Key10270631
MDR Text Key199299152
Report Number3003790304-2020-00051
Device Sequence Number1
Product Code FFK
Combination Product (y/n)N
PMA/PMN Number
K120303
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 08/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCW-USLTA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2020
Was the Report Sent to FDA? No
Date Manufacturer Received07/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-