MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PIN LNR CONS NEUT +4 36IDX56OD; PINNACLE HIP SYSTEM : HIP CONSTRAINED ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PIN LNR CONS NEUT +4 36IDX56OD; PINNACLE HIP SYSTEM : HIP CONSTRAINED ACETABULAR LINERS Back to Search Results
Model Number 1218-36-656
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Joint Dislocation (2374); No Code Available (3191)
Date of Event 07/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that patient had a right hip revision done somewhere in (b)(6) about 5 years ago.The surgeon put in a constrained liner because the patient was dislocating.The patient showed up to the er with a dislocated hip and surgeon said the patient had dislocated with a constrained liner from the x-ray and he wanted to do a new head/liner exchange.During the surgery, the surgeon removed the metal head and notice that there was metal debris, and the surgeon thought that this patient might have been dislocated for awhile because i guess this patient isn't very mobile.The surgeon removed the liner and implanted the same size constrained liner and chose to do a ceramic head of the same length.He was happy with the stability and leg lengths doi: 5 years ago; dor: (b)(6) 2020; affected side: right hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIN LNR CONS NEUT +4 36IDX56OD
Common Device Name
PINNACLE HIP SYSTEM : HIP CONSTRAINED ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10271155
Report Number1818910-2020-15786
Device Sequence Number3398998
Product Code KWZ
UDI-Device Identifier10603295012122
UDI-Public10603295012122
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K043058
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1218-36-656
Device Catalogue Number121836656
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/01/2020
Supplement Date Received by Manufacturer08/11/2020
Initial Report FDA Received Date07/14/2020
Supplement Report FDA Received Date08/12/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTICULEZE M HEAD 36MM +8.5; ARTICULEZE M HEAD 36MM +8.5
Outcome Attributed to Adverse Event Required Intervention;
Patient Age46 YR
-
-