DEPUY FRANCE SAS - 3003895575 CORAIL2 LAT COXA VARA SIZE 9; CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM Back to Search Results
Model Number 3L93709
Medical Device Problem Codes Loss of Osseointegration (2408); Unintended Movement (3026); Migration (4003)
Health Effect - Clinical Codes Adhesion(s) (1695); Pain (1994); Inadequate Osseointegration (2646); No Code Available (3191)
Date of Event 06/10/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Clinical notification received for revision of left total hip to address aseptic loosening and migration of the femoral stem.Date of implantation: (b)(6) 2009, date of revision: (b)(6) 2020, (left hip).Treatment: stem, head, and liner were revised.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Revision operative notes 06/10/2020 indicate the patient received a left total hip revision due to pain, with loosening, subsidence and varus angulation of the femoral stem.Upon entering the joint, the femoral stem was noted to be grossly loose.Significant scar tissue and bone was removed from the medial greater trochanter for access.Scar tissue was also noted to be removed from the acetabular side.The surgery was completed without indication of complication by the surgeon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL2 LAT COXA VARA SIZE 9
Common Device Name
CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
MDR Report Key10271730
Report Number1818910-2020-15803
Device Sequence Number14257437
Product Code KWA
UDI-Device Identifier10603295168867
UDI-Public10603295168867
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K042992
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 07/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3L93709
Device Catalogue Number3L93709
Device Lot Number2789111
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/01/2020
Supplement Date Received by Manufacturer07/15/2020
07/30/2020
08/10/2020
Initial Report FDA Received Date07/14/2020
Supplement Report FDA Received Date07/31/2020
08/04/2020
08/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CORAIL2 LAT COXA VARA SIZE 9; DLT TS CER HD 12/14 40MM +5.0; PINNACLE 100 ACET CUP 56MM; PINNACLE MTL INS NEUT40IDX56OD; CORAIL2 LAT COXA VARA SIZE 9; DLT TS CER HD 12/14 40 MM +5.0; PINNACLE 100 ACET CUP 56 MM; PINNACLE MTL INS NEUT40IDX56OD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
Patient Weight91
-
-