| Model Number |
AVSM10100 |
| Medical Device Problem Codes |
Break (1069); Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lot number was provided, a lot history review was performed.The device was returned to the manufacturer for evaluation.The investigation is confirmed for outer sheath fracture which made a successful deployment impossible.The definite root cause could not be determined based on the available information.The device is labeled for single use.
|
| |
|
Event or Problem Description
|
|
This report summarizes one malfunction.A review of the reported information indicated that model avsm10100 covera vascular covered stent system allegedly experienced broken sheath and failed to deploy.This information was received from one source.One patient was involved with no reported patient injury.The male patient is (b)(6) years old and weighs (b)(6) lbs.
|
| |
|
Search Alerts/Recalls
|