|
Other, other text: h3: two pictures of an inflation valve were provided for evaluation.The inflation valve appeared to be correctly sealed into the inflation balloon.The outer edge of the inflation valve appeared to be deformed on one side.One bivona custom tracheostomy tube was returned for evaluation.Visual inspection of the returned device confirmed that the inflation valve was seated correctly and that the portion of the valve that extends outside the balloon was deformed.The cuff on the device was unable to be inflated using the inflation valve in its current state since the tips of various sized syringes were unable to be inserted to actuate the valve.In order to inflate the cuff so the symmetry and inflation system could be checked, the deformed inflation valve was removed and another one was inserted into the balloon.The device inflated without issue.The device was leak tested and the cuff symmetry was measured.No leaks were detected for the entire device (e.G., balloon, airway line, neck flange, cuff, or lumen).Using a verified ruler (blue dot) the cuff symmetry passed the manufacturing specification of 1/3:2/3.No defects were identified with the device except for the deformity on the inflation valve.While the cuff was inflated, the non-deformed inflation valve was removed, and the original inflation valve was reinserted into the device.The inflation valve functioned and did not allow the fluid in the cuff or the balloon to leak.The investigation determined that it is unlikely that the deformity in the inflation valve was present when the device was released for distribution since all custom devices are 200% inspected for cuff symmetry and leaks prior to packaging and the inflation valve in its current state prevented the cuff from being inflated.Moreover, the complaint description details indicate that the device was able to be inflated.The investigation could not confirm whether the deformity in the inflation valve was caused by exposure to heat during reprocessing or force.
|