MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
|
Back to Search Results |
|
Model Number 97714 |
Device Problems
Malposition of Device (2616); Impedance Problem (2950); Insufficient Information (3190)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/13/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: product id: 97714, serial#: (b)(4), product type: implantable neurostimulator.Product id: neu_unknown_lead.Serial#: unknown.Product type: lead.Other relevant device(s) are: product id: neu_unknown_lead, serial/lot #: unknown.Refer to regulatory report (b)(4).The patient had two implanted systems and it was not clear which issues pertained to which implanted system.The referenced report pertains to the patient¿s other system.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from the manufacturer's representative regarding a patient who was implanted with a neurostimulator for spinal cord stimulation.It was reported that the impedance values on the two sub-q leads were not usable for therapy based on the impedance test.Rep stated the leads were trashed.
|
|
Manufacturer Narrative
|
Continuation of d11: product id 97714 lot# serial# (b)(6) implanted: (b)(6) 2016 explanted: product type implantable neurostim ulator product id 3888-45 lot# v302548 serial# implanted: (b)(6) 2009 explanted: product type lead medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received from the rep indicated that the issue was not resolved.They indicated that a new trial needs to be set.The rep noted that if the trial is successful, then full implant would occur and explant of the sub-q leads.However, with covid-19, they are unsure when the patient¿s trial will be, as everything is backed up for 6 or more months.
|
|
Manufacturer Narrative
|
Correction: continuation of d11: neurostimulator, product id: 3888-45, lot# v302548, implanted: (b)(6) 2009, product type lead.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was reported that the cause of the lead being trashed was due to it being placed off label sub-q.The patient's programming was moved to the remaining epidural lead.The issue has not been resolved, but the doctor is planning on trailing a new lead in the epidural space.
|
|
Manufacturer Narrative
|
Correction to h6: device code c63114 applies to this event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Continuation of d11: product id 97714 lot# serial# (b)(6) implanted: 2016(b)(6) explanted: product type implantable neurostim ulator product id 3888-45 lot# v302548 serial# implanted: 2009(b)(6) explanted: product type lead medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received from a manufacturer representative (rep).The rep reported that a trial date had nor been set yet.Covid-19 was making it difficult to get scheduled.
|
|
Search Alerts/Recalls
|
|
|