MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 2.25/22; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 2.25/22; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 364499
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/23/2020
Type of Reportable Event Malfunction
Event or Problem Description
An orsiro drug-eluting stent system was selected for treatment.As it was noted that the stent was not visible on the angiography the device was removed and there was no stent on the balloon.
 
Additional Manufacturer Narrative
The returned product was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation of the returned instrument revealed that the balloon has been inflated and was deflated in the as-returned condition.Microscopic analysis revealed stent imprints on the exposed balloon surface, indicating that the stent was initially crimped in between the two radiopaque markers.The stent was not returned.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.Based on the conducted investigations, no manufacturing or material related root cause could be determined.It should be noted that the ifu advises the user to verify that the stent is positioned between the proximal and distal balloon markers during stent system preparation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 2.25/22
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key10273524
Report Number1028232-2020-02915
Device Sequence Number16987014
Product Code NIQ
UDI-Device Identifier07640130439085
UDI-Public7640130439085
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/27/2021
Device Model Number364499
Device Catalogue NumberSEE MODEL NO.
Device Lot Number07195707
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/03/2020
Initial Date Received by Manufacturer 07/14/2020
Supplement Date Received by Manufacturer08/10/2020
Initial Report FDA Received Date07/14/2020
Supplement Report FDA Received Date08/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-