| Model Number |
10601 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/07/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is a combination product.
|
| |
|
Event or Problem Description
|
|
It was reported that stent damage occurred.A 3.00 x 12 synergy ii drug-eluting stent was selected for use.However, during preparation, it was found out that the stent edge was deformed on balloon.The procedure was completed with a different device.There were no patient complications nor injuries reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: synergy ii us mr 3.00 x 12 mm stent delivery system was returned for analysis.An examination (visual and via scope) of the crimped stent found stent damage.Stent struts from the distal region of the stent were noted to be lifted from their crimped position and pulled proximally.The undamaged crimped stent outer diameter was measured and the result was within max crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube shaft found no issues.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that stent damage occurred.A 3.00 x 12 synergy ii drug-eluting stent was selected for use.However, during preparation, it was found out that the stent edge was deformed on balloon.The procedure was completed with a different device.There were no patient complications nor injuries reported.
|
| |
|