| Model Number |
101740000 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/21/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The ac inlet solder joint has been damaged by excessive mechanical stress exerted on the joint via the ac cord either by multiple plugging/unplugging or mechanical stress applied to the ac cord.
|
| |
|
Event or Problem Description
|
|
Customer reported that they saw smoke coming out of the dc adapter.They immediately unplugged the cord and did not use it again.There were no injuries or other damage reported.
|
| |
|
Search Alerts/Recalls
|