| Model Number |
1910-5 |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
Low Oxygen Saturation (2477)
|
| Date of Event |
06/08/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device history record for lot m19135t505 was reviewed and the product was produced according to product specifications.The device were provided by the customer.All information reasonably known as of (b)(6) 2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical inc.Complaint database and identified as complaint (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the "non-return valve" [irrigation port cover] was missing from the device.Patient desaturated to 95% after five minutes.The device was replaced allowing the normalization of the oxygen saturation.Additional information received 29-jun-2020 indicated "the nurse quickly notice[d] the dysfunction, so there was no harmful consequence for the patient.".
|
| |
|
Additional Manufacturer Narrative
|
|
Photos of the device were provided by the customer.The photos were reviewed; however, no root cause could be determined.All information reasonably known as of 24 oct 2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
| |
|
Search Alerts/Recalls
|