| Catalog Number |
CAT00778 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported there was increased traction time.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: leaking adaptor.Probable root cause: design: poor interconnection design.Manufacturing: adapter or cannula not manufactured to specification.Application: user error - cannula/adapter not connected properly.Improper handling causes damage to cannula/adapter.The reported failure mode will be monitored for future reoccurrence.Manufacturing date is not known.H3 other text : 81.
|
| |
|
Event or Problem Description
|
|
It was reported there was increased traction time.
|
| |
|