• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ASSURA (DDQP+); IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ASSURA (DDQP+); IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3265-40
Medical Device Problem Code Over-Sensing (1438)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/24/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.Further information was requested but not received.
 
Event or Problem Description
It was reported that the patient presented remotely via merlin.Net.Review of transmissions revealed that the implantable cardioverter defibrillator exhibited post-paced t-wave oversensing.The patient did not receive therapy during the oversensing.No device intervention was performed but programming changes were discussed.No patient symptoms were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA (DDQP+)
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key10277098
Report Number2017865-2020-09180
Device Sequence Number2842142
Product Code NIK
UDI-Device Identifier05414734505666
UDI-Public05414734505666
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 07/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2014
Device Model NumberCD3265-40
Device Catalogue NumberCD3265-40
Device Lot Number3867989
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2020
Initial Report FDA Received Date07/15/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age55 YR
-
-