ADVANCED BIONICS, LLC HIRES 90K ADVANTAGE IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1500-04
Device Problem Migration (4003)
Patient Problems Itching Sensation (1943); Skin Irritation (2076); Swelling (2091)
Event Type  Injury  
Event Description
The recipient reportedly experienced a head trauma incident.The recipient presented with persistent swelling, redness, itchiness, and device migration.Programming adjustments were made, however, the issue did not resolve.The recipient's swelling changed with and without device use.The recipient's device was explanted.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The recipient was reimplanted with another advanced bionics cochlear device.Additional details will not be provided.The external visual inspection revealed the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Excessive swelling was noted during explant surgery.The recipient healed from surgery.The recipient will be reimplanted.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K ADVANTAGE IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key10279412
MDR Text Key199039630
Report Number3006556115-2020-00603
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016828638
UDI-Public(01)07630016828638(11)140930(17)170831
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup,Followup
Report Date 06/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date08/31/2017
Device Model NumberCI-1500-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-