Model Number CQ75124
Medical Device Problem Codes Material Frayed (1262); Inflation Problem (1310); Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review was performed.The sample was returned to the manufacturer for evaluation.Therefore, the investigation of the reported event is confirmed for the frayed material, identified longitudinal balloon rupture and unconfirmed for the reported inflation issue.Based upon the available information, the definitive root cause for this event is unknown.The device is labeled for single use.
 
Event or Problem Description
This report summarizes one malfunction.A review of the reported information indicated that model cq75124 pta balloon dilatation catheter allegedly experienced inflation issue, frayed and longitudinal rupture.This information was received from one source.The malfunction had patient involvement but the patient had no consequences.The patient was a (b)(6) year old female weighing (b)(6) lbs.
 
Event or Problem Description
This report summarizes one malfunction.A review of the reported information indicated that model cq75124 pta balloon dilatation catheter allegedly experienced inflation issue, frayed and longitudinal rupture.This information was received from one source.The malfunction had patient involvement but the patient had no consequences.The patient was a 62 year old female weighing 185 lbs.
 
Additional Manufacturer Narrative
H10: as the lot number for the device was provided, a manufacturing review was performed.The sample was returned to the manufacturer for evaluation.Therefore, the investigation of the reported event is confirmed for the frayed material, identified longitudinal balloon rupture and unconfirmed for the reported inflation issue.Based upon the available information, the definitive root cause for this event is unknown.The device is labeled for single use.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key10279442
Report Number2020394-2020-04465
Device Sequence Number19464388
Product Code DQY
UDI-Device Identifier00801741063251
UDI-Public(01)00801741063251
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberCQ75124
Device Catalogue NumberCQ75124
Device Lot NumberREDW0248
Initial Date Received by Manufacturer 06/30/2020
Supplement Date Received by Manufacturer06/30/2020
Initial Report FDA Received Date07/15/2020
Supplement Report FDA Received Date07/20/2020
Patient Sequence Number1
-
-