• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. REFLEXION SPIRAL VARIABLE RADIUS CATHETER 5 F LOOP BI-DIRECTIONAL, 7F; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. REFLEXION SPIRAL VARIABLE RADIUS CATHETER 5 F LOOP BI-DIRECTIONAL, 7F; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number D402865
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Cardiac Perforation (2513)
Date of Event 07/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.Further information regarding the event was requested but not received.
 
Event or Problem Description
Related manufacturer reference: 3008452825-2020-00345, 3005334138-2020-00304, 3008452825-2020-00353.During an atrial fibrillation ablation procedure, a pericardial effusion occurred.Following ablation of the pulmonary vein, the patient became hypotensive.An echocardiogram confirmed a pericardial effusion.The location and cause of the effusion remains unknown.Protamine was administered and no further intervention was required.There were no further complications and the patient was stabilized.There were no performance issues with any abbott device.
 
Additional Manufacturer Narrative
Additional information: g4, g7, h2, h3, h6, h10.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported pericardial effusion could not be conclusively determined.Per the ifu, cardiac perforation is a known risk during the use of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFLEXION SPIRAL VARIABLE RADIUS CATHETER 5 F LOOP BI-DIRECTIONAL, 7F
Common Device Name
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
14901 deveau place
minnetonka MN 55345
MDR Report Key10279816
Report Number2182269-2020-00066
Device Sequence Number14106739
Product Code DRF
UDI-Device Identifier05414734206211
UDI-Public05414734206211
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K072012
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Model NumberD402865
Device Catalogue NumberD402865
Device Lot Number7312999
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/01/2020
Supplement Date Received by Manufacturer07/21/2020
Initial Report FDA Received Date07/15/2020
Supplement Report FDA Received Date08/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BRK¿ XS TRANSSEPTAL NEEDLE.; SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER.; TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SE.
Outcome Attributed to Adverse Event Other;
-
-