| Catalog Number |
801-90038 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
06/11/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Visual inspection: the device was inspected and found to have a deformed distal tip.One of the four clamping jaws was bent outward which is consistent with the reported event details.It was also found that the primary swiveling component of the gold lever was galling.Signs of wear could be seen around the component.Functional inspection: the device was tested by depressing both of the levers and checking for issues.A considerable amount of resistance could be felt while depressing the gold lever which indicates that galling has occurred.Device and complaint history records were reviewed and no relevant manufacturing issues or similar complaints were identified.Since the instrument was manufactured over 5.5 years ago (and was over three (3) years old at the time of failure), the issue can most likely be attributed to routine wear/tear and cleaning/reprocessing of the device.Routine wear/tear and cleaning/reprocessing of the device appears to have worn out the original lubrication which led to galling and eventual device failure.When the instrument was locked intraoperatively, the swiveling component was most likely frozen due to the combination of insufficient lubrication, progressive galling, and high locking force.As a result, the failure mode was confirmed and the most likely root cause for the issue was determined to have been normal wear/tear and cleaning/reprocessing of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that a mesa dual action rod reducer dragonfly tip became deformed and was not able to be removed after intra-operative locking.The surgeon removed a portion of the patient's bone surrounding the instrument in an attempt to remove it.The surgeon was finally able to remove the instrument with a mallet and the procedure was completed successfully with a 10-20 minute surgical delay.
|
| |
|
Event or Problem Description
|
|
It was reported that a mesa dual action rod reducer dragonfly tip became deformed and was not able to be removed after intra-operative locking.The surgeon removed a portion of the patient's bone surrounding the instrument in an attempt to remove it.The surgeon was finally able to remove the instrument with a mallet and the procedure was completed successfully with a 10-20 minute surgical delay.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection: the device was inspected and found to have a deformed distal tip.One (1) of the four (4) clamping jaws was bent outward which is consistent with the reported event details.It was also found that the primary swiveling component of the gold lever was galling.Signs of wear could be seen around the component.Functional inspection: the device was tested by depressing both of the levers and checking for issues.A considerable amount of resistance could be felt while depressing the gold lever which indicates that galling has occurred.Device and complaint history records were reviewed and no relevant manufacturing issues or similar complaints were identified.Since the instrument was manufactured over 5.5 years ago, and was over three (3) years old at the time of failure, the issue can most likely be attributed to routine wear/tear and cleaning/reprocessing of the device.Routine wear/tear and cleaning/reprocessing of the device appears to have worn out the original lubrication which led to galling and eventual device failure.When the instrument was locked intraoperatively, the swiveling component was most likely frozen due to the combination of insufficient lubrication, progressive galling, and high locking force.As a result, the failure mode was confirmed and the most likely root cause for the issue was determined to have been normal wear/tear and cleaning/reprocessing of the device.
|
| |
|
Search Alerts/Recalls
|