SMITH & NEPHEW, INC. LENS 4K A/C CAMERA HEAD; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 72205058
Device Problems Optical Discoloration (2999); Optical Problem (3001)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/24/2020
Event Type  malfunction  
Event Description
It was reported that during knee scope procedure, the camera head had a blinking picture and distorted coloring.No delay and a back up device was used to complete the procedure.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The reported device, used in treatment, was returned to the designated complaint station for independent evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed and no deficiencies were observed.A functional test revealed video flickering when the cable near the camera connector was moved.The complaint was verified and the root cause was associated with component failure.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found related failures; this failure mode will be trended to assess for any necessary corrective actions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LENS 4K A/C CAMERA HEAD
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key10280722
MDR Text Key199057476
Report Number1643264-2020-00576
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier00885556720790
UDI-Public00885556720790
Combination Product (y/n)N
PMA/PMN Number
K191177
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72205058
Device Catalogue Number72205058
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2020
Date Manufacturer Received08/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-