Catalog Number XXX-HAKIM PROGRAMMABLE VALVE
Device Problem Positioning Problem (3009)
Patient Problem Injury (2348)
Event Type  Injury  
Manufacturer Narrative
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported the hakim valve accidentally reset during an abdominal intervention.After the surgery, the patient started to show symptoms of low brain drainage flow.A codman vpv programmer was used to set the valve to the original setting, resulting in patient recovery.
 
Manufacturer Narrative
The hakim valve was not returned for evaluation and lot number information has not been provided; therefore, an evaluation of the device could not be performed, and manufacturing records could not be reviewed.The cause(s) of the difficulty reported by the customer could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.The possible root cause for the programming issue reported by the customer could be due strong magnet field as noted in the ifu : subjecting the valve to strong magnetic fields may change the setting of the valve.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAKIM PROGRAMMABLE VALVE
Type of Device
CHPV
Manufacturer (Section D)
RAYNHAM
325 paramount drive
325 paramount drive
raynham MA 02767
MDR Report Key10280939
MDR Text Key201580523
Report Number1226348-2020-00311
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 05/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberXXX-HAKIM PROGRAMMABLE VALVE
Was Device Available for Evaluation? No
Date Manufacturer Received06/25/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-