| Model Number |
MMT-723RNAP |
| Medical Device Problem Codes |
No Display/Image (1183); Obstruction of Flow (2423); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/22/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that they insulin pump had blank display.No harm requiring medical intervention was reported.Customer received no delivery alarm but then did not turn back on.Customer stated that insulin pump was not exposed to moisture.The customer was assisted with troubleshooting and customer reports display was blank currently.Customer stated the display was not blank less than 30 seconds and did not return.The customer removes the battery and stated the insert battery alarm did not display on the insulin pump screen.Customer stated the contact on the battery cap was neither missing nor damage, only the coin slot of the battery cap was worn out.Customer stated battery compartment was neither damaged nor corroded.Customer stated the spring was neither damaged nor corroded.Customer stated that display did not returned after insulin pump restart.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device power up properly after battery installation.No blank display or display anomaly noted.However, device alarmed failed battery test due to corroded battery tube.Unable to perform operating currents, self-test, a21 error, rewind test, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and displacement test or verify no delivery alarm due to failed battery test alarm.Unable to verify battery cap damage due to battery cap not come with unit.(b)(4).
|
| |
|