| Model Number |
MMT-723LNAS |
| Medical Device Problem Codes |
Mechanical Problem (1384); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/10/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump drive support cap was flushed.The insulin pump had motor error alarm.Customer was able to successfully clear the alarm.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test, rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery test.No motor error alarm noted.Motor passed motor test.The drive support disk was inspected and no anomaly noted.(b)(4).
|
| |
|