Model Number WCD 4000
Device Problem Reset Problem (3019)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/17/2020
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of monitor (b)(4) has been completed.  the reported problem (resets) has been confirmed.  upon investigation the monitor reset during incoming testing.  the cause for the failure was isolated to shorted insulated-gate bipolar transistors (igbts) q12 and q16 on the defibrillator pca.The root cause for the shorted igtbs could not be positively identified.No adverse event resulted from the defective monitor.
 
Event Description
A us distributor reported that a monitor reset.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh, pa
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh, pa
Manufacturer Contact
rachel mahoney
121 gamma drive
pittsburgh, pa 
9683333
MDR Report Key10283757
MDR Text Key199677853
Report Number3008642652-2020-06059
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005005
UDI-Public00855778005005
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 07/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Device Catalogue Number10A0988
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/17/2020
Initial Date Manufacturer Received 07/09/2020
Initial Date FDA Received07/16/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/29/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-