MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AUTOTOME RX 49; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AUTOTOME RX 49; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number M00545160
Medical Device Problem Codes Material Integrity Problem (2978); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/17/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).(b)(4).The returned autotome rx 49 was analyzed, and a visual evaluation noted that the device did not present any visual damage in the tip or the cutting wire.However, the working length was twisted.The reported event was not confirmed.The failure found could had been generated during manipulation of the device or due to the interaction with the endoscope or with other devices.Based on the investigation the most probable cause for working length twisted is adverse event related to procedure since the adverse event occurred during the procedure and the device had no influence on event.Additionally, all compiled information on this investigation determines that the most probable cause for the reported complaint is no problem detected since the device problem (tip cracked) cannot be confirmed.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the directions for use (dfu) / product label.
 
Event or Problem Description
It was reported to boston scientific corporation that an autotome rx 49 was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2020.According to the complainant, during the procedure, the physician noted a visible crack at the tip of the device.The procedure was not completed as the patient could not endure the prolonged procedure.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOTOME RX 49
Common Device Name
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key10284028
Report Number3005099803-2020-02817
Device Sequence Number14169332
Product Code KNS
UDI-Device Identifier08714729415152
UDI-Public08714729415152
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K013153
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/11/2022
Device Model NumberM00545160
Device Catalogue Number4516
Device Lot Number0023639962
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/01/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/18/2020
Initial Report FDA Received Date07/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age43 YR
Patient Weight47
-
-