Model Number 71335552
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Arthralgia (2355)
Event Date 06/30/2020
Event Type  Injury  
Event Description
It was reported that, after a tha had been performed, the patient experienced pain.The adverse event was addressed by performing a revision surgery.The patient outcome is unknown.
 
Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical evaluation concluded that based on the limited information, the root cause of the ¿pain¿ which precipitated the revision cannot definitively be concluded.The patient¿s stated rheumatoid arthritis and poor bone stock might be the major source of the pain experienced by this patient and the poor bone stock could be factor for the revision and not a failure of the components.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use documents revealed this failure mode and potential harm was previously identified.Potential probable causes could be but not limited to traumatic injury, joint tightness, material in use, patient reaction, design of device, bone quality, and patient health post-operatively.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved or product information, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 3 HOLE ACET SHELL 52MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10284271
MDR Text Key199119351
Report Number1020279-2020-03272
Device Sequence Number1
Product Code MBL
UDI-Device Identifier03596010598240
UDI-Public03596010598240
Combination Product (y/n)N
PMA/PMN Number
K070756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 06/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/16/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71335552
Device Catalogue Number71335552
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Date Manufacturer Received06/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
REF THREADED HOLE COVER (CN 71336500); REF THREADED HOLE COVER (CN 71336500)
Patient Outcome(s) Hospitalization; Required Intervention;
-
-