Model Number 1550300-12
Medical Device Problem Codes Material Rupture (1546); Improper or Incorrect Procedure or Method (2017); Difficult to Advance (2920)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/23/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer reported the device is not returning.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat a heavily calcified lesion in the moderately tortuous left anterior descending (lad) coronary artery.A rotablator was used and balloon angioplasty was performed.The 3.00x12 mm xience sierra stent delivery system (sds) failed to cross so pre-dilatation was performed with a cutting balloon.The xience sierra sds was reinserted and the stent was implanted at 16 atmospheres.Upon removal, blood was noted inside the balloon, indicating a rupture.There were no adverse patient effects and there was no reported clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no similar incidents from this lot.It was reported that the device was reinserted into the patient.It should be noted that the xience sierra, everolimus eluting coronary stent system, instructions for use (ifu) states: an unexpanded stent may be retracted into the guiding catheter one time only.An unexpanded stent should not be reintroduced into the artery once it has been pulled back into the guiding catheter.Subsequent movement in and out through the distal end of the guiding catheter should not be performed, as the stent may be damaged or dislodge during retraction back into the guiding catheter.It is unknown if the ifu deviation directly caused or contributed to the reported event.The investigation determined the reported difficult to advance and material rupture appears to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10284928
Report Number2024168-2020-05900
Device Sequence Number2679002
Product Code NIQ
UDI-Device Identifier08717648227240
UDI-Public08717648227240
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/10/2021
Device Model Number1550300-12
Device Catalogue Number1550300-12
Device Lot Number808304A
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/23/2020
Supplement Date Received by Manufacturer07/24/2020
Initial Report FDA Received Date07/16/2020
Supplement Report FDA Received Date08/12/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-