Model Number RMU-1000 |
Device Problem
Failure to Run on Battery (1466)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Although requested, the device associated with the report has not yet be returned.The investigation is ongoing and a root cause is not currently known.Should additional information become available a follow-up mdr will be submitted.
|
|
Event Description
|
A fire department reported that their chest compressor stopped compressions during a rescue attempt.It was also reported that the patient was not resuscitated.No additional event or device information was provided.Although requested, the device associated with this event has not been returned and the cause for the event is not known.It is not known if the device was removed and if manual compression could be performed; therefore, this mdr is being submitted out of an abundance of caution.
|
|
Manufacturer Narrative
|
Analysis of the device identified that the rmu-1000's top screw became loose causing the reported device problems.Although the top screw came loose, this incident does not appear to be a production issue as the device's motor screw was torqued and a loctite patch was cured as required by the manufacturing process specification - mps.
|
|