MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. DORADO PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. DORADO PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number DR13572
Medical Device Problem Codes Break (1069); Deflation Problem (1149); Inflation Problem (1310)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lot number for the device was provided, a manufacturing review was performed.The device was returned to bd and is pending evaluation.The company is still investigating the issue at this time.
 
Event or Problem Description
This report summarizes one malfunction.A review of the reported information indicated that the model dr13572 pta dilatation catheter allegedly failed to inflate and deflate.This information was received from one source.The malfunction involved a patient with no known impact to the patient.The patient's age, sex and weight were not reported.
 
Additional Manufacturer Narrative
H10: the lot number for the device was provided, and a lot history review was performed.The device was returned to the manufacturer for inspection/evaluation.The investigation is confirmed for inflation problem and break; however, the investigation is inconclusive for deflation problem.Based upon the available information, the definitive root cause for this event is unknown.The device is labeled for single use.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
This report summarizes one malfunction.A review of the reported information indicated that the model dr13572 pta dilatation catheter allegedly failed to inflate and deflate.This information was received from one source.The malfunction involved a patient with no known impact to the patient.The patient's age, sex and weight were not reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORADO PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key10286691
Report Number2020394-2020-04530
Device Sequence Number485716
Product Code DQY
UDI-Device Identifier00801741058936
UDI-Public(01)00801741058936
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberDR13572
Device Catalogue NumberDR13572
Device Lot Number93XD0110
Initial Date Received by Manufacturer 06/30/2020
Supplement Date Received by Manufacturer09/30/2020
Initial Report FDA Received Date07/16/2020
Supplement Report FDA Received Date10/13/2020
Patient Sequence Number1
-
-