| Model Number |
CQ7568 |
| Medical Device Problem Codes |
Inflation Problem (1310); Peeled/Delaminated (1454); Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a device history review was performed.The sample was returned to the manufacturer for inspection/evaluation.Investigation of the received sample confirmed the alleged malfunction.Balloon rupture was noted to be the likely root cause for the inflation issue.However, the definitive root cause for the balloon rupture and peeled pebax could not be determined based on available information.It is unknown whether patient or procedural issues contributed to the event.The device is labeled for single use.
|
| |
|
Event or Problem Description
|
|
This report summarizes one malfunction.A review of the reported information indicated that model cq7568 pta dilatation catheter allegedly experienced pinhole balloon rupture and had inflation issue.This information was received from one source.Of the one malfunction, one patient was involved with no patient consequence or impact.The patient was reported to be a (b)(6) year old female weighing (b)(6) lbs.
|
| |
|